EasyManua.ls Logo

Fluke 712B/EN - General Specifications

Fluke 712B/EN
38 pages
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
712B
Users Manual
32
General Specifications
Maximum voltage applied between any
terminal and earth ground or between
any two terminals:
30 V
Operating temperature -10 °C to 50 °C
Storage temperature -20 °C to 60 °C
Operating altitude 2,000 meters
Storage altitude 12,000 meters
Relative Humidity (% RH operating
without condensation)
Non condensing
90 % (10 °C to 30 °C)
75 % (30 °C to 40 °C)
45 % (40 °C to 50 °C)
(Without condensation)
Vibration Requirements MIL-T-28800E, Class 2
Drop Test Requirements 1 meter
IP Rating IEC 60529: IP52
Electromagnetic Environment IEC 61326-1, Portable
Safety IEC 61010-1, Max 30 V to earth, Pollution Degree 2
Power Supply 4 AA alkaline batteries/NEDA code: 15A, IEC code: LR6
Size (H x W x L) 52.5 x 84 x 188.5 mm
Weight 524 g
w ww . . co m
information@itm.com1.800.561.8187

Related product manuals