RTD/Thermocouple Calibrator
714B Specifications
S
-20 to 0 2.3 2.8 1.2 1.8
0 to 200 1.5 2.1 1.1 1.7
200 to 1400 0.9 1.4 0.9 1.4
1400 to 1767 1.1 1.7 1.0 1.5
C
0 to 800 0.6 0.9 0.6 0.9
800 to 1200 0.8 1.2 0.7 1.0
1200 to 1800 1.1 1.6 0.9 1.4
1800 to 2316 2.0 3.0 1.3 2.0
L
-200 to -100 0.6 0.9 0.3 0.4
-100 to 800 0.3 0.4 0.2 0.3
800 to 900 0.5 0.8 0.2 0.3
U
-200 to 0 0.6 0.9. 0.4 0.6
0 to 600 0.3 0.4 0.3 0.4
BP
0 to 1000 1.0 1.5 0.4 0.6
1000 to 2000 1.6 2.4 0.6 0.9
2000 to 2500 2.0 3.0 0.8 1.2
XK
-200 to 300 0.2 0.3 0.2 0.5
300 to 800 0.4 0.6 0.3 0.6
G
100 to 300 1.6 2.4 1.2 1.8
300 to 1500 1.0 1.5 1.0 1.5
1500 to 2315 2.0 3.0 1.6 2.4
D
0 to 300 1.6 2.4 1.2 1.8
300 to 1500 1.0 1.5 1.0 1.5
1500 to 2315 2.0 3.0 1.6 2.4
P
0 to 1000 1.6 2.4
0.6 0.9
1000 to 1395 2.0 3.0 0.8 1.2
M
-50 to 100 1.0 1.5 0.4 0.6
100 to 1000 1.6 2.4 0.6 0.9
1000 to 1410 2.0 3.0 0.8 1.2
Notes:
Sensor inaccuracies not included.
Accuracy with external cold junction; for internal junction add 0.2 °C
Temperature scale: ITS-90
Compensation: NIST Monograph 175 for B, R, S, E, J, K, N, T. DIN 43710 for L, U. GOST P 8.585-2001 (Russia) for
BP and XK. ASTM E988-96 for C. ASTM E1751/E1751M 09ԑ1 for G, D, P, M
Resolution: 0.1 °C
Temperature Coefficient: 0.05 °C/°C ( <18 °C or >28 °C )
0.07 °C/°C for C type >1800 °C and for BP type >2000 °C
7