EasyManua.ls Logo

Fluke 712B - Verify 4-Wire Ohms Measure

Fluke 712B
50 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
712B/714B
Calibration Manual
Table 4. 712B Ohms Source Limits
1 Yr. Lower 1 Yr. Upper 2 Yr. Lower 2 Yr. Upper
8508A Range Source ()
Limit () Limit () Limit () Limit ()
1 0.950 1.050 0.920 1.080 2k range, 4W
10 9.949 10.052 9.917 10.083 2k range, 4W
100 99.935 100.065 99.890 100.110 2k range, 4W
390 389.892 390.109 389.803 390.197 2k range, 4W
1000 999.35 1000.65 998.90 1001.10 20k range, 4W
4000 3998.90 4001.10 3998.00 4002.00 20k range, 4W
Verify 4-Wire Ohms Measure
1. Push to set the 712B to Ohms Measure mode.
2. Push until 4W shows on the screen.
3. Make the connections between the 712B and the 5522A, as shown in
Figure 3.
4. Set the 5522A to output the Ohms value at the second column in Table 5.
5. Wait approximately 10 seconds for a stable output of the 5522A, and verify
that the 712B readings are within
the limits shown.
Fluke 5522A
5522A
CALIBRATOR
Fluke 712B
hwp003.eps
Figure 3. Verify 712B 4-Wire Ohms Measure
12

Table of Contents

Other manuals for Fluke 712B

Related product manuals