RTD/Thermocouple Calibrator
714B Calibration Adjustment
Adjust mA Measure
1. Connect the 5522A and the 714B via the Fluke-75X-8014 cable, as shown in
Figure 19.
The 5522A outputs 0.1 mA.
hwp26.jpg
Fluke 5522A
5522A
CALIBRATOR
Fluke 714B
hwp023.eps
Figure 19.714B and 5522A mA Channel Connections