712B/714B
Calibration Manual
General Specifications
Maximum voltage applied between
any terminal and earth ground or
between any two terminals:
30 V
Operating temperature
-10 °C to 50 °C
Storage temperature
-20 °C to 60 °C
Operating altitude 2,000 meters
Storage altitude 12,000 meters
Relative Humidity (% RH operating
without condensation)
Non condensing
90 % (10 °C to 30 °C)
75 % (30 °C to 40 °C)
45 % (40 °C to 50 °C)
(Without condensation)
Vibration Requirements MIL-T-28800E, Class 2
Drop Test Requirements 1 meter
IP Rating IEC 60529: IP52 (with TC cap)
Electromagnetic Environment
IEC 61326-1, Portable
FCC: CFR Title 47, Part 15, Subpart B
Safety IEC 61010-1, Max 30 V to earth, Pollution Degree 2
Electromagnetic Compatibility
Applies to use in Korea only. Class A Equipment (Industrial Broadcasting &
Communication Equipment)
This product meets requirements for industrial (Class A) electromagnetic
wave equipment and the seller or user should take notice of it. This
equipment is intended for use in business environments and is not to be
used in homes.
Power Supply 4 AA alkaline batteries/NEDA code: 15A, IEC code: LR6
Size (H x W x L) 52.5 x 84 x 188.5 mm
Weight 515 g