EasyManua.ls Logo

Fluke 712B - DC mA and Ohms Measurement Specifications; DC mA Measurement; Ohms Measurement

Fluke 712B
50 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
712B/714B
Calibration Manual
Adjust mA Measure
Before the calibration adjustment, set the 5522A to mA (dc) source output mode,
and use one set of banana plug patch cords to connect the 712B and the 5522A,
as shown in Figure 13.
Fluke 5522A
5522A
CALIBRATOR
Fluke 712B
hwp022.eps
Figure 13. Adjust 712B mA Measure
Adjust 0.1 mA Input
1. Set the 5522A to output 0.1 mA.
2. Push on the 712B to enter 0.1 mA input calibration mode.
Wait approximately 10 seconds for a stable output of the 5522A.
3. Confirm the reading on the 712B.
hwp14.jpg
28

Table of Contents

Other manuals for Fluke 712B

Related product manuals