712B/714B
Calibration Manual
Adjust 50 Input
1. Set the 5522A to output 50 .
2. Wait approximately 10 seconds for a stable output of the 5522A.
3. Confirm the reading on the 712B.
hwp10.jpg
Adjust 350 Input
1. Push on the 712B to enter 350 input calibration mode.
2. Set the 5522A to output 350 .
Wait approximately 10 seconds for a stable output of the 5522A.
3. Confirm the reading on the 712B.
hwp11.jpg