RTD/Thermocouple Calibrator
712B Calibration Adjustment
Adjust 24 mA Input
1. Set the 5522A to output 24 mA.
2. Push on the 712B to enter 24 mA input calibration mode.
Wait approximately 10 seconds for a stable output of the 5522A.
3. Confirm the reading on the 712B.
hwp15.jpg
Adjust External Trigger for Ohms Source
Before the calibration, set the 5522A to mA (dc) source output mode, and use
one set of banana plug patch cords to connect the 712B and the 5522A, as
shown in Figure 14.
Fluke 5522A
5522A
CALIBRATOR
Fluke 712B
hwp005.eps
Figure 14. Adjust External Trigger for Ohms Source