718 30G/100G
Users Manual
14
Using an External Pump
W Warning
To avoid damage to the Calibrator and possible
release of pressure, do not connect the internal
sensor to an external pressure source that
exceeds 30 psi for Model 718 30G or 100 psi for
Model 718 100G.
To develop higher pressure or vacuum, use an external
pump (such as the Fluke Model 700PTP). Use a Fluke
Pressure Module connected to the pressure module input
on the Calibrator. Pressure modules are listed in Table .
Make overall connections as shown in Figure 6.
Refer to setup and operating instructions included with the
pressure module and pump.