EasyManua.ls Logo

Fluke 725 - Page 17

Fluke 725
18 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Multifunction Process Calibrator
Specifications
15
Temperature, RTD Ranges, and Accuracies
(ITS-90)
Accuracy
Type Range °C Measure
4-Wire °C
Measure
2- and
3-Wire*
°C
Source °C
Ni120 -80 to 260 0.2 0.3 0.2
Pt100-385 - 200 to 800 0.33 0.5 0.33
Pt100-392 -200 to 630 0.3 0.5 0.3
Pt100-JIS -200 to 630 0.3 0.5 0.3
Pt200-385 -200 to 250
250 to 630
0.2
0.8
0.3
1.6
0.2
0.8
Pt500-385 -200 to 500
500 to 630
0.3
0.4
0.6
0.9
0.3
0.4
Pt1000-385 -200 to 100
100 to 630
0.2
0.2
0.4
0.5
0.2
0.2
Resolution: 0.1 °C, 0.1 °F
Allowable excitation current (source): Ni120, Pt100-385,
Pt100-392, Pt100-JIS: 0.15 to 3.0 mA
Pt200-385, Pt500-385: 0.05 to 0.80 mA; Pt1000-385: 0.05
to 0.40 mA
RTD Source: Addresses pulsed transmitters and PLCs
with pulses as short as 5 ms.
* 2-wire: Does not include lead resistance.
3-wire: Assumes matched leads with a total resistance
not exceeding 100 .
General Specifications
Operating
temperature
-10 °C to 55 °C
Storage temperature - 20 °C to 71 °C
Operating altitude
3000 meters above mean sea
level
Relative Humidity (%
RH operating without
condensation)
90 % (10 to 30 °C)
75 % (30 to 40 °C)
45 % (40 to 50 °C)
35 % (50 to 55 °C)
uncontrolled < 10 °C
Vibration Random, 2 g, 5 to 500 Hz
Safety
EN 61010-1:1993, ANSI/ISA
S82.01-1994; CAN/CSA C22.2
No 1010.1:1992
Power requirements 4 AA alkaline batteries
Size
96 x 200 x 47 mm. (3.75 x 7.9
x 1.86 in)
Weight 650 gm (1 lb, 7 oz)
Protection Class Pollution Degree II

Other manuals for Fluke 725