725
Product Overview
14
Temperature, Thermocouples
Type
Range
Measure and Source
Accuracies (ITS-90)
J -200 to 0 °C
0 to 1200 °C
1.0 °C
0.7 °C
K -200 to 0 °C
0 to 1370 °C
1.2 °C
0.8 °C
T -200 to 0 °C
0 to 400 °C
1.2 °C
0.8 °C
E -200 to 0 °C
0 to 950 °C
0.9 °C
0.7 °C
R -20 to 0 °C
0 to 500 °C
500 to 1750 °C
2.5 °C
1.8 °C
1.4 °C
S -20 to 0 °C
0 to 500 °C
500 to 1750 °C
2.5 °C
1.8 °C
1.5 °C
B 600 to 800 °C
800 to 1000 °C
1000 to 1800 °C
2.2 °C
1.8 °C
1.4 °C
L -200 to 0 °C
0 to 900 °C
0.85 °C
0.7 °C
U -200 to 0 °C
0 to 400 °C
1.1 °C
0.75 °C
N -200 to 0 °C
0 to 1300 °C
1.5 °C
0.9 °C
XK -200 to -100 °C
-100 to 800 °C
0.5 °C
0.6 °C
BP 0 to 800 °C
800 to 2500 °C
1.2 °C
2.5 °C
Resolution:
J, K, T, E, L, N, U, XK, BP: 0.1 °C, 0.1 °F
B, R, S: 1 °C, 1 °F
Loop Power Supply
Voltage: 24 V
Maximum current: 22 mA
Short circuit protected
RTD Excitation (simulation)
Allowable Excitation by RTD type
Ni 120 0.15 to 3.0 mA
Pt 100-385 0.15 to 3.0 mA
Pt 100-392 0.15 to 3.0 mA
Pt 100-JIS 0.15 to 3.0 mA
Pt 200-385 0.05 to 0.80 mA
Pt 500-385 0.05 to 0.80 mA
Pt 1000-385 0.05 to 0.40 mA
Pressure Measurement
Range Resolution Accuracy Units
Determined
by pressure
module
5 digits Determined
by pressure
module
psi, inH
2
O@4 °C,
inH
2
O@20 °C, kPa,
cmH
2
O@4 °C,
cmH
2
O@20 °C, bar,
mbar, kg/cm
2
,
mmHg, inHg