Testing an Output Device
45
Testing an Output Device
Use the source functions to test and calibrate actuators,
recording, and indicating devices. Proceed as follows:
1. Connect the test leads to the instrument under test
as shown in Figure 22.
2. Press
V for current or dc voltage, or F for
frequency or resistance (lower display).
3. If necessary, press
M for SOURCE mode.
V mA
LOO P
V mA
TC RTD
˚
C
˚
F
Hz
100%
25%
25%
RECALL
ZERO
MEAS
SOURCE
STORE
SETUP
0%
MULTIFUNCTION CALIBRATOR
725
Black
Red
0 to 1 V dc
Input
sh25f.eps
Figure 22. Calibrating a Chart Recorder