Specifications
5
DC Voltage Specifications, Output
Ranges
[1]
Absolute Uncertainty, tcal ±5°C,
±(ppm of output +µV)
Stability
Resolution
Maximum
Burden
[2]
90 Days 1 Year
24 hours, ±1 °C
±(ppm of output +μV)
0 mV to 100.000 mV 25 3 30 3
5 ppm + 2 μV 1 μV
10 mA
0 V to 1.00000 V 25 10 30 10
4 ppm + 10 μV 10 μV
10 mA
0 V to 10.0000 V 25 100 30 100
4 ppm + 100 μV 100 μV
10 mA
0 V to 100.000 V 25 1 mV 30 1 mV 5 ppm + 1 mV 1 mV 1 mA
[1] All outputs are positive only, unless otherwise noted.
[2] Remote sensing is not provided. Output resistance is < 1Ω.
Noise
Ranges
Bandwidth 0.1 to 10 Hz p-p
±(ppm of output +µV)
Bandwidth 10 Hz to
10 kHz rms µV
0 mV to 100.000 mV
1 μV 6 μV
0 V to 1.00000 V
10 μV 60 μV
0 V to 10.0000 V
100 μV 600 μV
0 V to 100.000 V 10 ppm+1 mV 20 mV
DC Voltage Specifications, Thermocouple Jack, Input and Output
Ranges
[1]
Absolute Uncertainty, tcal ±5°C,
±(ppm of output +µV)
Stability
Resolution
Maximum
Burden
[2]
90 Days 1 Year
24 hours, ±1 °C
±(ppm of output +μV)
-10 to 75.000 mV 25
2 μV
30
2 μV 5 ppm + 2 μV 1 μV 10 Ω
[1] All outputs are positive only, unless otherwise noted.
[2] Remote sensing is not provided. Output resistance is < 1Ω.
DC Voltage Specifications, Isolated Input
Ranges
Absolute Uncertainty, tcal ±5 °C,
±(ppm of reading + mV)
Resolution
0 V to 10.0000 V 50 0.2
100 μV
0 V to 100.000 V 50 2.0 1 mV
DC Current Specifications, Output
Ranges
[1]
Absolute Uncertainty, tcal ±5 °C,
± (ppm of output + µA)
Resolution
Maximum
Compliance
Voltage
Maximum
Inductive
Load
90 Days 1 Year
0 mA to 100.000 mA 40
[2]
1 50
[2]
1
1 μA
12 V 100 mH
[1] All outputs are positive only.
[2] For line voltages less than 95 V (±100 ppm of reading)
Ranges
Noise
Bandwidth 0.1 to 10 Hz p-p
Bandwidth 10 Hz to 10 kHz rms μV
0 mA to 100.000 mA 2000 nA
20 μA
DC Current Specifications, Isolated Input
Ranges
Absolute Uncertainty, tcal ±5 °C,
±(ppm of reading + μA)
Resolution
0 mA to 50.0000 mA 100 1
0.1 μA
0 mA to 24.0000 mA (Loop Power)
[1][2]
100 1
0.1 μA
[1] Loop Power: 24 V ±10 %
[2] HART Resistor: 250 Ω ±3 %