Precision Process Calibrator
Performance Tests
13
DC Voltage
The dc voltage performance test verifies the accuracy of the Volts Source output from
the front panel of the Product. Verify the measurements shown in Table 4 with a digital
multimeter (DMM).
Table 4. DC Voltage Verification
Range Nominal Value Measured Value
90-Day
Specification
(V)
1-Year
Specification
(V)
100 mV
0 V 3.00E-06 3.00E-06
0.035 V 3.88E-06 4.05E-06
0.065 V 4.63E-06 4.95E-06
0.1 V 5.50E-06 6.00E-06
1.0 V
0 V 1.00E-05 1.00E-05
0.35 V 1.88E-05 2.05E-05
0.65 V 2.63E-05 2.95E-05
1 V 3.50E-05 4.00E-05
10.0 V
0 V 1.00E-04 1.00E-04
3.5 V 1.88E-04 2.05E-04
6.5 V 2.63E-04 2.95E-04
10 V 3.50E-04 4.00E-04
100.0 V
0 V 1.00E-03 1.00E-03
35 V 1.88E-03 2.05E-03
65 V 2.63E-03 2.95E-03
100 V 3.50E-03 4.00E-03