Precision Process Calibrator
Performance Tests
17
Thermocouple Input
The TC input performance test verifies the accuracy of the TC input. Push and then
on the Product to set the Product to TC input. All Product conditions will be the
same as the Thermocouple Output test mode (CJC off, mV/ °C). Set the 5522A to the mV
values listed in Table 6. See Figure 5 for connections.
CURRENT
OUTPUT
HI
VOLTS
mA
RTD/
OUTPUT
HI
LO
HI
LO
LO
TC
SENSE
INPUT/OUTPUT
4
W RTD/
INPUT
20V PK
MAX
100 V MAX
100 mA M A X
20V PK
MAX
20V PK
MAX
PRECISION PROCESS CALIBRATOR
7526A
5522A
CALIBRATOR
HI
7526A
Calibrator
gxm013.eps
Figure 5. TC Input Performance Test Connections