753/754
Users Manual
40
To measure temperature where an RTD input is used:
1. If necessary, push for MEASURE mode.
2. Push t.
3. Push and then Select RTD Type is shown.
4. Push or to select the necessary RTD type.
5. Push .
6. Push or to select a 2-, 3-, or 4-wire connection.
The connections are shown on the display.
7. Attach the RTD to input jacks as the display or Figure
14 shows. Use the supplied jumper between the mA Ω
RTD MEASURE low jack and the V MEASURE low jack
as shown if using a 3-wire connection.
8. Push .
Caution
To prevent possible damage to the Product, do
not force a dual banana plug between any two
jacks in the horizontal orientation. Doing so
will damage the jacks. Use the supplied jumper
wire when needed for RTD measurements. A
dual banana plug may be used in the vertical
orientation. See Figure 16.
1.800.868.7495info@Fluke-Direct.ca