Specifications
Current Ranges ±20.99 mA ±21.0 mA -±99.9 mA
Resolution 0.01 mA 0.1 mA
Accuracy
20.99 mA range
99.9 mA range
0.2 % reading ±5 digits
1 % reading ±5 digits
Maximum Reading ±99.9 mA
Influence of Earth’s Field < 0.20 mA
Battery 2 AA 1.5 V Alkaline, IEC LR6
Working hours 45 hours
Size (H X W X L)
59 mm x 38 mm x 212 mm
(with clamp nested)
Weight
260g (Including battery)
Operating Temperature -10 to 50 °C
Storage Temperature -25 to 70 °C
Operating Humidity < 90 % @ <30 °C,
<75 % @ 30 to 50 °C
Operating Altitude 0 to 2000 m
Storage Altitude None
IP Rating IP 40
Vibration Requirements Random 2 g, 5 to 500 Hz
EMI, RFI, EMC Meets all applicable requirements in
EN 61326-1
Temperature Coefficients
0.1x(specified accuracy)/°C
(< 18 °C or > 28 °C)
Measurement Category
IEC 61010-1
61010-2-032
CAT II 300 V
CAT II Equipment is designed to
protect against transients from energy-
consuming equipment supplied from
the fixed installation, such as TVs,
PCs, portable tools, and other
household appliances.
Agency Approvals
P, ), ;, FCC
N10140

Other manuals for Fluke 771