772/773
Calibration Manual
10
Volts DC Measure Accuracy Tests
UUT Meter Reading Limit
Step Calibrator Output
Low High
1 0.00 V -0.02 V 0.02 V
2 10.00 V 9.96 V 10.04 V
3 -10.00 V -10.04 V -9.96 V
4 20.00 V 19.94 V 20.06 V
5 -20.00 V -20.06 V -19.94 V
6 30.00 V 29.92 V 30.08 V
7 -30.00 V -30.08 V -29.92 V
mA DC Source Accuracy Tests
DMM Reading Limit
Step UUT Meter Output
Low High
1 0.00 mA -0.02 mA 0.02 mA
2 4.00 mA 3.97 mA 4.03 mA
3 8.00 mA 7.96 mA 8.04 mA
4 12.00 mA 11.96 mA 12.04 mA
5 20.00 mA 19.94 mA 20.06 mA
6 24.00 mA 23.93 mA 24.07 mA
Volts DC Source Accuracy Tests
DMM Reading Limit
Step UUT Meter Output
Low High
1 0.00 V -0.02 V 0.02 V
2 2.50 V 2.47 V 2.53 V
3 5.00 V 4.97 V 5.03 V
4 7.50 V 7.46 V 7.54 V
5 10.00 V 9.96 V 10.04 V
Calibration Adjustment
The Meter features closed-case calibration adjustment with a known reference source.
The Meter measures the applied reference source, calculates correction factors, and keeps
the correction factors in nonvolatile memory.
Before you start calibration adjustment, let the Meter stabilize to room temperature.