8600A
Section
4
Maintenance
WARNING!
THESE
SERVICING
INSTRUCTIONS
ARE
FOR
USE BY
QUALIFIED
PERSONNEL
ONLY.
TO
AVOID
ELECTRIC
SHOCK,
DO
NOT
PERFORM
ANY
SERVICING
OTHER
THAN
THAT
CONTAINED
IN
THE
OPERATING
INSTRUCTIONS
UNLESS
YOU ARE
QUALIFIED
TO DO
SO.
4-1.
INTRODUCTION
4-2.
This
section
of
the manual
contains
information
concerning
maintenance
and servicing
of
the Model
8600A
Digital
Multimeter.
A calibration
interval
of
6
months
is
recommended
to
insure
instrument
operation
within
the
6 month
specifications.
Test
equipment
recommended
for
performance
tests,
calibration
adjustments
and trouble-
shooting
is listed
in Table
4-1
.
If
the
recommended
equip-
ment
is not
available,
equipment
of
equivalent
specifi-
cations
may
be
used.
4-3.
SERVICE
INFORMATION
44.
Each
instrument
manufactured
by the John Fluke
Mfg.
Co., Inc.
is warranted
for
a period
of
one year
upon de-
livery
to the
original
purchaser.
The WARRANTY
is printed
on
the
back of the
title
page
located
at the front
of the
man-
ual.
4-5.
Factory
authorized
calibration
and service
for each
Fluke product
is
available
at various
world wide
locations.
A
complete
list
of these
Technical
Centers
is
included
in
Section
7.
Shipping
information
is given
in
Section
2 of this
manual. If
requested,
an
estimate
will
be provided
to the
customer before
any
repair
work
is
begun
on
instruments
that
are beyond
the
warranty
period.
Table 4-1.
REQUIRED
TEST EQUIPMENT
NOMENCLATURE
MINIMUM
USE
SPECIFICATIONS
RECOMMENDED
EQUIPMENT
AC
Calibrator
Voltage
Range:
Oto
1000V ac
Frequency
Range:
30Hz
to
100
kHz
Voltage
Accuracy:
30 Hz
to 50 Hz
0.05%
50 Hz
to 10 kHz
0.02%
10 kHz
to 100 kHz.
.
.
0.05%
John Fluke
Model
5200A
and
5205A
DC
Calibrator
Voltage
Range:
Oto 100V
dc
Accuracy:
±0.003%
John
Fluke
Model
343A
DC
Current
Current
Range:
0 to 2 mA
John
Fluke
Calibrator
Accuracy:
0.006%
Model
38JA
DC
Current
Current
Range:
2 mA
to 2A
John Fluke
Calibrator
Accuracy:
0.02%
Model
382A
Digital
Multimeter
Voltage
Accuracy:
0.1%
Input
Impedance:
1000MJ2
John Fluke
Model
8600A
Resistor
Decade
Resistance
Accuracy:
+0.01%
ESI
1063B
Oscilloscope
General
Purpose
with 10
Mfi
probe
Tektronix
465
1/77
4-1