6-19. Introduction
6-20.
The Model 801-600, as shown in Figure
6-6,
is a
clamp-on
current
probe which is used to extend the current
measurement
capabilities of
the
8600A.
The probe
is
designed to measure
currents of 2 to 600 amperes at fre-
quencies of up to 400 Hz
with
±3%
accuracy. The clamp-
on
feature allows current to be
measured without breaking
the
circuit
under
test.
6-21. Operation
6-22. Use the
following procedure for
operating
the
8600A
with the 801-600
probe:
a.
Plug the
801-600 dual-banana plug into
the MA and
COMMON
INPUT
terminals on
the 8600A.
Figure
6-6.
AC
CURRENT PROBE,
CLAMP-ON
(801-600)
6-4