EasyManua.ls Logo

Fluke 8600A - 4-24. CALIBRATION

Fluke 8600A
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8600A
Table
4-3.
OHMS
PERFORMANCE
TEST
8600A
RANGE
INPUT
REQUIRED
8600A DISPLAY
LIMITS
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
SHORT
100.00ft
1
.OOOOkft
lO.OOOkft
lOO.OOkft
lOOO.Okft
lO.OOOMft
OPEN
00.00
to
00.03
99.87
to 100.13
.9989
to
1.0011
9.994
to
10.006
99.94
to
100.06
999.4
to 1000.6
9.979
to
10.021
18.888
Flashing
(20Mft
LED
lit)
4-20.
DC
Current
Test
4-2
1 . Using
the
dc
current
calibrator
(check
Table
4-
1
for
correct
calibrator)
sequentially
apply
the
values
of
dc
current
indicated
in
Table
44
to
the
8600A MA
input
terminals
and
select
the
ranges
prescribed.
The
8600A
should
display
a
reading
within
the
indicated
limits.
Table
44.
DC
CURRENT
PERFORMANCE
TEST
RANGE
INPUT
READING
200
/r
A
200
mA
2 mA
20
mA
200 mA
2000
mA
Open
+1
90.00 pA
+1
.9000
mA
+19.000
mA
+190.00
mA
+1
.9000
A
-00.05
to +00.05
+189.79
to
+190.21
+1.8979
to +1.9021
+18.979
to
+19.021
+189.79
to
+190.21
+1897.9
to
+1902.1
NOTE!
Because
the
current
measurements
of
ac and
dc
are
made
using
the same
shunt
resistors;
a check
of
ac
current
is
not
made.
4-22.
AC
Voltage
T
est
4-23.
Using
the
ac
voltage
calibrator,
apply
the ac vol-
tages
indicated
in
Table
4-5
to
the
8600A V-f2
input
ter-
minals
and
select
the
ranges
prescribed.
The
8600A should
display
a reading
within
the indicated
limits.
4-24.
CALIBRATION
4-25.
Introduction
4-26
The
8600A should
be
calibrated
every
6 months
or
whenever
repairs
have been
made,
to
insure
the
instrument
continues
to
operate
at its
rated
accuracy.
After
calibration
is
completed,
the
DC
current
performance
test,
paragraph
4-20
should
be
run
through
to
verify
the
functioning
and
accuracy
of the
current
dividers.
The
calibration
should
be
performed
under
the
following
environmental
conditons;
ambient
temperature
of
23°C
±5°C
and
a relative
humidity
of
less
than
80%.
Refer
to
Table
4-1
for
the
recommended
test
equipment.
Calibration
adjustment
locations
are
pictured
in
Figure
4-1.
Perform
the
follow-
ing
preliminary
steps
before
calibrating
the
instrument.
a.
Remove
the
instrument
from
the
case.
b.
Connect
the
8600A
to
the
appropriate
(1
1 5V
ac or
230V
ac)
line
power.
c.
Turn
the
instrument
on
and
allow
it to
warm-up
for
a
minimum
of
15
minutes.
d.
Insure
that
the
ac
and
dc
calibrators
are
up to
their
normal
operating
temperatures.
NOTE!
Use
only
non-metalic
adjustment
tools.
Table 4-5.
ACV
PERFORMANCE
TEST
8600A
RANGE
INPUT
REQUIRED
FRE-
QUENCY
Hz
8600A
DISPLAY
LIMITS
200
mV
190.00
mV
70kHz
188.05
to 191.95
mV
200
mV
190.00
mV
500
Hz
189.46
to 190.54
mV
200
mV
39.00
mV
30
kHz
38.60 to
39.40 mV
2 V
1.9000
V
500
Hz
1.8959
to
1.9041
V
2
V
.9000 V
500
Hz
.8979
to
.9021
V
2
V
1
.9000
V
50 kHz
1.8900
to 1.9100
V
2 V
1.9000
V
100
kHz
1.8800
to
1.9200
V
2 V
1.0000
V
30
Hz
.9945
to 1.0055 V
20
V
19.000
V
500
Hz
18.959
to
19.041
V
20
V
19.000
V
70 kHz
18.800
to 19.200
V
200
V
190.00
V
500 Hz
189.59
to 190.41 V
200
V
100.00
V
30 kHz
99.45
to 100.55 V
1200
V
1000.0
V
500 Hz
996.9
to 1003.1 V
1200
V
1000.0
V
20 kHz
994.0
to
1006.0
V
4-27.
DC
Volts
Calibration
4-28
. The
calibration
procedure
for
the
DCV
FUNCTION
of
the
8600A is
presented
in
Table
4-6.
Use
the
recommended
dc
voltage
calibrator
(see
Table
4-1)
to
apply
the
prescribed
dc
voltages
to the V-f2
INPUT
terminals
and,
where
required,
make
the
adjustments
to
meet
the
specified
display
limits.
1/77
4-3

Table of Contents

Related product manuals