8600A
Table
6-3.
8600A RANGES
FOR
DC HV
PROBE (80K-40)
RANGE
VOLTAGE
RANGE
SELECTED WITH
80K-40
PROBE
200
20 to
40
kV
20
2 to 20 kV
2
1 to 2 kV
6-29.
HIGH
FREQUENCY PROBE (80RF-1)
6-30.
Introduction
6-31
. The
Model 80RF-1 High
Frequency Probe,
Figure
6-9,
extends
the
frequency range
of the 8600A to
include
100
kHz to 500
MHz for
ac
voltage
measurements
from
0.25
to 30V rms. The 80RF-1
operates in conjunc-
tion with the dc voltage
ranges, and is connected
to the
8600A using
a
shielded
dual-banana plug and an
adapter.
6-32.
Specifications
Voltage:
0.25V to
30V
Response : Responds to
peak value of
input.
Calibrated
to
read
rms
value of a sine
wave
input.
AC to DC
Transfer
Accuracy:
Loaded
with
10 megohms
±
10
%.
6-33. Operating Notes
6-34.
The
straight
and the hooked tips supplied with the
probe are useful for making
voltage measurements up
to
100
MHz. For measurements
above 100
MHz use the
high
frequency
adapter tip with
mating
connector and 50
ohm terminations.
6-35.
The maximum input to the probe is
30V
rms
or
200V dc.
These voltage limits
may be used in
combination
so
that
the ac
component of an
ac
signal
superimposed on
dc level can be measured.
CAUTION
Changing the dc level of the input signal by
more
than 200
volts will damage
the
probe.
100 kHz- 100
MHz-
100
MHz 500
MHz
+10°C to +30°C
+5%
+7%
—
10°C
to
+40°C
+7%
+15%
< ±3
db at 10 kHz and
700 MHz
Input Impedance: 4 megohms shunted by 2
±0.5
pf
Maximum Input: 30 volts rms ac, 200
volts
dc
Cable Connections: Shielded dual banana
plug
fits
all standard %-inch dual
banana connectors.
Cable Length : 4
ft.
(1
2
1 .9
cm)
minimum
Weight:
3
l
A
oz.net
Accessories: Ground lead, straight tip
,
hook tip, high
frequency
adapter
Figure
6-9. 80RF-1.HIGH FREQUENCY
PROBE
6-36. Operation
6-37.
Use the following procedure for
operating the
8600A with
the
80RF-1
probe:
a. Connect the 80RF-1 shielded
dual-banana
plug
to the 8600A
V-Sl
and LO
INPUT terminals.
b.
Attach
the desired
probe tip
to the probe
body.
c.
Depress the
DCV
pushbutton
(FUNCTION)
d.
Select the
20,
200,
or 1200V range.
NOTE
The probe loading error caused
by
the
100
OMNI
input
impedqnce on the
8600A
200
mVand
2V
ranges can
be corrected
by
placing a
10
MSI
±5%
resistor
across
the
dual-banana plug
terminals.
6-6