87
Users Manual
38
Specifications (cont)
Function Range Resolution Accuracy***
Capacitance 5.00 nF
0.0500 µF
0.500 µF
5.00 µF
0.01 nF
0.0001 µF
0.001 µF
0.01 µF
±(1% + 3)
±(1% + 3)
±(1% + 3)
±(1.9% + 3)
Diode Test 3.000V 0.001V ±(2% + 1)
Function Range Resolution Accuracy* Burden Voltage (typical)
mA †
\
(45 Hz to 2 kHz)
mA
[
40.00 mA
400.0 mA
4000 mA
10.00A††
40.00 mA
400.0 mA
4000 mA
10.00 A††
0.01 mA
0.1 mA
1 mA
0.01A
0.01 mA
0.1 mA
1 mA
0.01A
±(1.0% + 2)
±(1.0% + 2)
±(1.0% + 2)
±(1.0% + 2)
±(0.2% + 2)
±(0.2% + 2)
±(0.2% + 2)
±(0.2% + 2)
1.8 mV/mA
1.8 mV/mA
0.03 V/A
0.03 V/A
1.8 mV/mA
1.8 mV/mA
0.03 V/A
0.03 V/A
*** With film capacitor or better, using Relative mode to zero residual.
† See page 34 for a complete explanation of this notation.
†† 10A continuous, 20A for 30 seconds maximum.W