Specifications
39
Specifications (cont)
Function Range Resolution Accuracy Burden Voltage (typical)
B
µA
(45 Hz to 2 kHz)
F
µA
400.0 µA
4000 µA
400.0 µA
4000 µA
0.1 µA
1 µA
0.1 µA
1 µA
±(1.0% + 2)
±(1.0% + 2)
±(0.2% + 3)
±(0.2% + 2)
100 µV/µA
100 µV/µA
100 µV/µA
100 µV/µA
Function Range Resolution Accuracy
Frequency
(0.5 Hz to 200 kHz, pulse
width >2 µs)
199.99
1999.9
19.999 kHz
199.99 kHz
>200 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 kHz
±(0.005% + 1)
±(0.005% + 1)
±(0.005% + 1)
±(0.005% + 1)
Unspecified
† See page 36 for a complete explanation of this notation.