Remote Operation
Error Codes
97
Table 33. Execution Error Codes and Descriptions (cont.)
Error Code Error Text Description
-230 Data stale A FETCh? command was received but internal reading
memory was empty. The reading retrieved may be invalid.
-242 Date timeout Timed out while taking data.
-243 Second function invalid Secondary function is not enabled.
-318 Can’t save MAC Could not save MAC address.
-319 Can’t save LAN
configuration
Could not save network configuration.
-320 Can’t save non-volatile
configuration
Could not save non-volatile configuration.
-330 Self-test failed The Meter’s complete self-test failed from the remote interface
(*TST? command). In addition to this error, more specific self-
test errors are also reported.
-350 Too many errors The error queue is full because more than 16 errors have
occurred. No additional errors are stored until you remove
errors from the queue. The error queue is cleared when power
has been off, or after a *CLS (clear status) command has been
executed.
-410 Query interrupted A command was received which sends data to the output
buffer, but the output buffer contained data from a previous
command (the previous data is not overwritten). The output
buffer is cleared when power has been off, or after a *RST
(reset) command has been executed.
-420 Query
UNTERMINATED
The Meter was addressed to talk (i.e., to send data over the
interface) but a command has not been received which sends
data to the output buffer. For example, a CONFigure command
(which does not generate a response) may have been
executed, followed by an attempt to read data from the remote
interface.
-430 Query DEADLOCKED A command was received that generated too much data to fit
in the output buffer, and the input buffer is also full. Command
execution continues but all data is lost.
-440 Query
UNTERMINATED after
indefinite response
The *IDN? command must be the last query command within a
command string. Example: *IDN?;:SYST:VERS?
-501 RTC Time Error setting the RTC/System time.
-502 RTC Data Error setting the RTC/System date.
31 mA AC/DC selected
while REAR inputs
selected.
8842A emulation error. The Front/Rear input switch was set to
REAR while ac/dc current was selected via a remote command
or ac/dc current was selected and the Front/Rear switch was
switched to REAR.