Loading...
8845A/8846A
Programmers Manual
98
Table 33. Execution Error Codes and Descriptions (cont.)
Error Code Error Text Description
32 Offset selected while
reading
unavailable/overrange
8842A emulation error. The B1 offset command was entered
when a measurement wasn't available (such as external
triggering without a previous measurement) or the previous
measurement resulted in an overload condition.
51 8840 CALibration
command not valid
8842A emulation error. A 8842A calibration command was
entered. The 8845A/46A does not support 8842A calibration
commands.
71 8840 Syntax Error 8842A emulation error. An improper command was entered.
This error will occur when the command header is not formed
properly and fails to pass 8842A header syntax rules.
200 IG Invalid Invalid response type from inguard.
201 IG Packet size Incorrect packet size from inguard.
202 Trigger ignored Trigger ignored.
203 Configuration
acknowledgement
Wrong number configuration acknowledgement.
222 Parameter invalid Invalid parameter.
223 FUNC2 Function/2nd func mismatch
224 Math mismatch Function/math mismatch
225 Range mismatch Function/range mismatch.
226 Load fail Measurement configuration load failed.
300 GX Sync send Inguard not responding (send).
301 GX Sync recv Inguard not responding (recv),
302 GX Sync error Lost sync with inguard.
303 GX Port fail Could not open guard crossing port.
310 GX Ack queue full Acknowledgement queue full.
311 GX Pack sequence Acknowledgement queue full.
320 GX Ack Guard crossing not acking packets.
321 GX Pack over Guard crossing oversize packet recieved.
322 GX Ack num Guard crossing wrong ack number.
325 GX Ack byte Guard crossing unknown ack byte.
330 GX CRC failed Guard crossing Bad CRC.
340 GX Info packet Guard crossing info packet recieved; link not active.
350 GX Control byte Guard crossing unknown control byte.
360 GX Quality low Guard crossing quality indicator too low.
370 GX Restart Guard crossing restarted.