Remote Operation
Error Codes
99
Table 33. Execution Error Codes and Descriptions (cont.)
Error Code Error Text Description
421 Instrument configuration
invalid
Tried to set invalid state.
422 Instrument configuration
parameter
Invalid parameter.
426 Instrument configuration
load
Instrument configuration load failed.
510 CCO Name bad CCO constant name is bad.
511 RS-232
framing/parity/overrun
error
512 RS-232 fatal error A fatal error occured opening the serial port.
513 RS-232 Configuration
error
A fatal error occured while configuring the serial port.
514 Command allowed only
with RS-232
Commands which are only allowed with the RS-232/Ethernet
interface are SYSTem:LOCal, SYSTem:REMote, and
SYSTem:RWLock, or in Fluke 45 emulation mode the
commands are LOCS, REMS, RWLS, and LWLS.
520 Command line too long The command line exceeded 350 characters.
521 Input buffer overflow
522 Output buffer overflow
531 Insufficient memory There is not enough memory to store the requested number of
readings in internal memory using the INITiate command. The
product of the sample count (SAMPle:COUNt) and the trigger
count (TRIGger:COUNt) must not exceed 5,000 readings.
550 Command not allowed
in local
The multimeter received a READ? command while in the local
mode. For RS-232 operation, the SYSTem:REMote command
should always be executed before sending other commands
over the interface.
1000 Unknown function Unknown Function Selection.
1001 Unknown range Unknown Range Selection.
1002 Unknown CCO Unknown Calibration Constant.
1003 Can’t set NPLC Illegal/Unknown NPLC Selection.
1004 Can’t set trigger Illegal/Unknown TRIGGER Selection.
1005 Can’t get DCV Gains Cannot get 1V/10V DC linearization constants.
1006 Line Freq High AC Line frequency too high.
1007 Line Freq Low AC Line frequency too low.
1008 Write CCO failure Calibration Constant write failed.
65535 Unknown Unknown Error.