914X Field Metrology Wells
Prog Menu
30
RUN. Press Enter and the instrument will engage and start the test within a few seconds. Exit to
the main screen to view the test progress, refer to the Menu Structure.
When the switch resets, the test completes and the values of the switch OPEN, switch CLOSE,
and switch BAND are displayed for the user to record. The values may also be recorded inter-
nally in the instrument by selecting the option to record the data (-P model only).
Figure 13 Auto and manual switch test operation example