9
Specications and Environmental Conditions
Specications
2 SpecicationsandEnvironmentalConditions
2.1 Specications
Table 2 Metrology Well Specications
Specication 9170 9171 9172 9173
Range –45°C to 140°C
(–49°F to 284°F)
–30°C to 155°C
(–22°F to 311°F)
35°C to 425°C
(95°F to 797°F)
50°C to 700°C
(122°F to 1292°F)
Display Accuracy
2
±0.1°C full range ±0.1°C: 35°C to 100°C
±0.15°C: 100°C to 225°C
±0.2°C: 225°C to 425°C
±0.2°C: 50°C to 425°C
±0.25°C: 425°C to 660°C
Stability
1
±0.005°C full range ±0.005°C: 35°C to 100°C
±0.008°C: 100°C to
225°C
±0.01°C: 225°C to 425°C
±0.005°C: 50°C to 100°C
±0.01°C: 100°C to 425°C
±0.03°C: 425°C to 700°C
AxialUniformity
2
(40mm[1.6in])
±0.08°C: –45°C to –35°C
±0.04°C: –35°C to 0°C
±0.02°C: 0°C to 50°C
±0.07°C: 50°C to 140°C
±0.025°C: –30°C to 0°C
±0.02°C: 0°C to 50°C
±0.05°C: 50°C to 155°C
±0.05°C: 35°C to 100°C
±0.09°C: 100°C to 225°C
±0.17°C: 225°C to 425°C
±0.09°C: 50°C to 100°C
±0.22°C: 100°C to 425°C
±0.35°C: 425°C to 700°C
AxialUniformity
2
(60mm[2.3in])
±0.1°C: –45°C to –35°C
±0.04°C: –35°C to 0°C
±0.02°C: 0°C to 50°C
±0.07°C: 50°C to 140°C
±0.025°C: –30°C to 0°C
±0.02°C: 0°C to 50°C
±0.07°C: 50°C to 155°C
±0.05°C: 35°C to 100°C
±0.1°C: 100°C to 225°C
±0.2°C: 225°C to 425°C
±0.1°C: 50°C to 100°C
±0.25°C: 100°C to 425°C
±0.4°C: 425°C to 700°C
AxialUniformity
2
(80mm[3.15in])
N/A ±0.05°C: –30°C to 0°C
±0.04°C: 0°C to 50°C
±0.15°C: 50°C to 155°C
±0.06°C: 35°C to 100°C
±0.12°C: 100°C to 225°C
±0.23°C: 225°C to 425°C
±0.15°C: 50°C to 100°C
±0.30°C: 100°C to 425°C
±0.45°C: 425°C to 700°C
RadialUniformity ±0.01°C full range ±0.01°C: 35°C to 100°C
±0.02°C: 100°C to 225°C
±0.025°C: 225°C to
425°C
±0.01°C: 50°C to 100°C
±0.025°C: 100°C to
425°C
±0.04°C: 425°C to 700°C
LoadingEffect
(withreference
thermometer)
±0.02°C: –45°C to –35°C
±0.005°C: –35°C to
100°C
±0.01°C: 100°C to 140°C
±0.005°C: –30°C to 0°C
±0.005°C: 0°C to 100°C
±0.01°C: 100°C to 155°C
±0.01°C full range ±0.02°C: 50°C to 425°C
±0.04°C: 425°C to 700°C
Hysteresis ±0.025°C ±0.04°C ±0.07°C
Operating Range1 5°C to 40°C (41°F to 104°F)
Well Depth 160 mm (6.3 in) 203 mm (8 in)
Resolution 0.001° C/F
Display LCD, °C or °F user–selectable
KeyPad Ten key with decimal and +/– key. Function keys, menu key, and °C/°F key.
CoolingTime 44 min: 23°C to –45°C
19 min: 23°C to –30°C
19 min: 140°C to 23°C
30 min: 23°C to –30°C
25 min: 155°C to 23°C
220 min: 425°C to 35°C
100 min: 425°C to 100°C
235 min: 700°C to 50°C
153 min: 700°C to 100°C
HeatingTime 32 min: 23°C to 140°C
45 min: –45°C to 140°C
44 min: 23°C to 155°C
56 min: –30°C to 155°C
27 min: 35°C to 425°C 46 min: 50°C to 700°C
Size 366 x 203 x 323 mm (14.4 x 8 x 12.7 in) [height x width x depth]
Weight 14.2 kg (31.5 lb) 14.6 kg (32 lb) 12.2 kg (27 lb) 14.2 kg (31 lb)

Other manuals for Fluke 917 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 917 Series and is the answer not in the manual?

Fluke 917 Series Specifications

General IconGeneral
BrandFluke
Model917 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals