EasyManua.ls Logo

Fluke 917 Series - Specifications and Environmental Conditions; Specifications

Fluke 917 Series
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9
Specications and Environmental Conditions
Specications
2 SpecicationsandEnvironmentalConditions
2.1 Specications
Table 2 Metrology Well Specications
Specication 9170 9171 9172 9173
Range –45°C to 140°C
(–49°F to 284°F)
–30°C to 155°C
(–22°F to 311°F)
35°C to 425°C
(95°F to 797°F)
50°C to 700°C
(122°F to 1292°F)
Display Accuracy
2
±0.1°C full range ±0.1°C: 35°C to 100°C
±0.15°C: 100°C to 225°C
±0.2°C: 225°C to 425°C
±0.2°C: 50°C to 425°C
±0.25°C: 425°C to 660°C
Stability
1
±0.005°C full range ±0.005°C: 35°C to 100°C
±0.008°C: 100°C to
225°C
±0.01°C: 225°C to 425°C
±0.005°C: 50°C to 100°C
±0.01°C: 100°C to 425°C
±0.03°C: 425°C to 700°C
AxialUniformity
2
(40mm[1.6in])
±0.08°C: –45°C to –35°C
±0.04°C: –35°C to 0°C
±0.02°C: 0°C to 50°C
±0.07°C: 50°C to 140°C
±0.025°C: –30°C to 0°C
±0.02°C: 0°C to 50°C
±0.05°C: 50°C to 155°C
±0.05°C: 35°C to 100°C
±0.09°C: 100°C to 225°C
±0.17°C: 225°C to 425°C
±0.09°C: 50°C to 100°C
±0.22°C: 100°C to 425°C
±0.35°C: 425°C to 700°C
AxialUniformity
2
(60mm[2.3in])
±0.1°C: –45°C to –35°C
±0.04°C: –35°C to 0°C
±0.02°C: 0°C to 50°C
±0.07°C: 50°C to 140°C
±0.025°C: –30°C to 0°C
±0.02°C: 0°C to 50°C
±0.07°C: 50°C to 155°C
±0.05°C: 35°C to 100°C
±0.1°C: 100°C to 225°C
±0.2°C: 225°C to 425°C
±0.1°C: 50°C to 100°C
±0.25°C: 100°C to 425°C
±0.4°C: 425°C to 700°C
AxialUniformity
2
(80mm[3.15in])
N/A ±0.05°C: –30°C to 0°C
±0.04°C: 0°C to 50°C
±0.15°C: 50°C to 155°C
±0.06°C: 35°C to 100°C
±0.12°C: 100°C to 225°C
±0.23°C: 225°C to 425°C
±0.15°C: 50°C to 100°C
±0.30°C: 100°C to 425°C
±0.45°C: 425°C to 700°C
RadialUniformity ±0.01°C full range ±0.01°C: 35°C to 100°C
±0.02°C: 100°C to 225°C
±0.025°C: 225°C to
425°C
±0.01°C: 50°C to 100°C
±0.025°C: 100°C to
425°C
±0.04°C: 425°C to 700°C
LoadingEffect
(withreference
thermometer)
±0.02°C: –45°C to –35°C
±0.005°C: –35°C to
100°C
±0.01°C: 100°C to 140°C
±0.005°C: –30°C to 0°C
±0.005°C: 0°C to 100°C
±0.01°C: 100°C to 155°C
±0.01°C full range ±0.02°C: 50°C to 425°C
±0.04°C: 425°C to 700°C
Hysteresis ±0.025°C ±0.04°C ±0.07°C
Operating Range1 5°C to 40°C (41°F to 104°F)
Well Depth 160 mm (6.3 in) 203 mm (8 in)
Resolution 0.001° C/F
Display LCD, °C or °F user–selectable
KeyPad Ten key with decimal and +/– key. Function keys, menu key, and °C/°F key.
CoolingTime 44 min: 23°C to –45°C
19 min: 23°C to –30°C
19 min: 140°C to 23°C
30 min: 23°C to –30°C
25 min: 155°C to 23°C
220 min: 425°C to 35°C
100 min: 425°C to 100°C
235 min: 700°C to 50°C
153 min: 700°C to 100°C
HeatingTime 32 min: 23°C to 140°C
45 min: –45°C to 140°C
44 min: 23°C to 155°C
56 min: –30°C to 155°C
27 min: 35°C to 425°C 46 min: 50°C to 700°C
Size 366 x 203 x 323 mm (14.4 x 8 x 12.7 in) [height x width x depth]
Weight 14.2 kg (31.5 lb) 14.6 kg (32 lb) 12.2 kg (27 lb) 14.2 kg (31 lb)

Other manuals for Fluke 917 Series

Related product manuals