917X Series Metrology Wells
Environmental Conditions
10
Specication 9170 9171 9172 9173
Power 115 V (±10%), 50/60 Hz, 550 W
230 V (±10%), 50/60 Hz, 550 W
115 V (±10%), 50/60 Hz, 1025 W
230 V (±10%), 50/60 Hz, 1025 W
SystemFuse
Ratings
115 V: 6.3 A 250 V
230 V: 3.15 A 250 V
115 V: 10 A 250 V
230 V: 5 A 250 V
Internal Fuse
Ratings
F1: 8 A 250 V
F2: 1 A 250 V
F4: 4 A 250 V
n/a
ComputerInterface RS-232 Interface and 9930 Interface-it control software included
Safety OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC–61010–1:2001
1
Specications are given with an ambient temperature of 23°C (73.4°F). Range, display accuracy, axial uniformity, loading effect, cooling time, and heating
time are subject to the ambient temperature and may be affected outside the “Full Accuracy” temperature range
2
Refer to the Maintenance section in the User’s Guide regarding maintaining the accuracy of the instrument.
Table 3 Built-in Readout Specications
Specications Built-in Reference Input
TemperatureRange
1
–200°C to 962°C
(–328°F to 1764°F)
Resistance Range
0 to 400, auto-ranging
Characterizations ITS-90 subranges 4, 6, 7, 8, 9, 10, and
11
Callendar-Van Dusen (CVD): R0,
ALPHA, DELTA, BETA
Resistance Accuracy
1
0 to 20: 0.0005
20 to 400: 25 ppm (0.0025%)
TemperatureAccuracy
1, 2
10 PRTs:
±0.013°C at 0°C
±0.013°C at 50°C
±0.014°C at 155°C
±0.014°C at 225°C
±0.019°C at 425°C
±0.028°C at 700°C
25 and 100 PRTs:
±0.005°C at –100°C
±0.007°C at 0°C
±0.011°C at 155°C
±0.013°C at 225°C
±0.019°C at 425°C
±0.027°C at 661°C
TemperatureResolution 0.001° C/F
Operating Range 5°C to 40°C (41°F to 104°F)
Calibration NIST-traceable calibration provided
RecommendedProbes 5626-15-D or 5614-12-D3
1
Specications are given with an ambient temperature of 23°C (73.4°F). Resistance accuracy and temperature accuracy are subject to the ambient
temperature and may be affected outside the “Full Accuracy” temperature range.
2
The temperature range may be limited by the reference probe connected to the external “Probe” connection of the Metrology Well. Does not include
sensor probe accuracy. It does not include probe uncertainty or probe characterization errors.
3
The 5614–12–D’s temperature range is 420°C.
2.2 EnvironmentalConditions
Although the instrument has been designed for optimum durability and trouble-free
operation, it must be handled with care. The instrument should not be operated in an
excessively dusty or dirty environment. Maintenance and cleaning recommendations
can be found in the Maintenance section.
The instrument operates safely under the following environmental conditions:

Other manuals for Fluke 917 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 917 Series and is the answer not in the manual?

Fluke 917 Series Specifications

General IconGeneral
BrandFluke
Model917 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals