EasyManuals Logo

Fluke 917 Series User Manual

Fluke 917 Series
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
15
Quick Start
Display
3.3 SystemSetup
Before using the instrument, the parameters in this section need to be setup (Main
Menu: SYSTEM|SETUP MENU).
3.4 Display
Language
Select the preferred language (English, French, or Chinese) using the right/left arrow
keys and press “ENTER” to accept the selection.
Decimal
The decimal of the numbers in the instrument can be either a comma or decimal.
Select the desired decimal type using the right/left arrow keys and press “ENTER” to
accept the selection.
3.4.1 Measure
Stability Limit
NOTE: Metrology Wells should not be expected to operate better than the
stability limit specication set forth in the Specications section of this guide.
Therefore, the minimum setting of the stability limit should not be less than the
stability specication set forth in the Specications section.
The stability limit of the instrument is the parameter which allows the instrument to
notify the user when it has achieved the stability limit set in this parameter. There are
two notications: visual and audible. The visual notication is always active. When
the instrument is operating within the stability limit, the stability parameter on the
main screen will remain highlighted as long as the instrument is within the given
specication, otherwise the parameter will not be highlighted. The audible, if enabled,
alerts the user once per set-point when the instrument achieves the set stability limit.
Use the numeric keys to set the desired stability limit and press “ENTER” to accept
the new stability limit.
Example:
A specic calibration process requires the instrument be operating within ±0.1°C.
“0.1” would be entered into the stability limit parameter. When the instrument is
within ±0.1°C, “STAB: X.XXX°C” will be highlighted and the audible alarm (if
enabled) will notify the user that the instrument is operating within ±0.1°C. Use the
numeric keys to set the desired stability limit and press “ENTER” to accept the new
stability limit.

Other manuals for Fluke 917 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 917 Series and is the answer not in the manual?

Fluke 917 Series Specifications

General IconGeneral
BrandFluke
Model917 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals