917X Series Metrology Wells
Setup
14
●
Getting Started Guide
●
User’s Guide
●
Report of Calibration and calibration label
●
DIN Connector (-R units only)
●
Tongs (insert removal tool)
9173
●
9173 Metrology
Well
●
9173-INSX Insert (X=A, B, C, D, E, or F)
●
Power Cord
●
RS-232 Cable
●
Getting Started Guide
●
User’s Guide
●
Report of Calibration and calibration label
●
DIN Connector (-R units only)
●
Tongs (insert removal tool)
If all items are not present, contact an Authorized Service Center.
3.2 Setup
NOTE: The instrument will not heat, cool, or control until the “CONT
ENABLE” parameter is set to “On.” Refer to Set-Point Setup on page 16 to set
this parameter.
Place the calibrator on a at surface with at least 6 inches of free space around the in-
strument. Overhead clearance is required. DO NOT place under a cabinet or structure.
Plug the Metrology Well power cord into a mains outlet of the proper voltage, fre-
quency, and current capability (see Specications for power details). Observe that the
nominal voltage corresponds to that indicated on the back of the calibrator.
Carefully insert the probe sleeve into the well. Probe sleeves should be of the small-
est hole diameter possible still allowing the probe to slide in and out easily. Sleeves
of various sizes are available from Fluke Calibration. The well must be clear of any
foreign objects, dirt and grit before the sleeve is inserted. The sleeve is inserted with
the two small tong holes positioned upward.
Turn on the power to the calibrator by toggling the switch on the power entry module.
After a brief self-test, the controller should begin normal operation. The main screen
will appear within 30 seconds. If the instrument fails to operate, please check the
power connection.
The display will show the well temperature, and wait for user input before heating or
cooling to current set-point.