DPM2 Plus
Operators Manual
2
Safety Information
W Caution
To avoid personal injury and possible damage to the
Meter:
Use the Meter only with gases and liquids found in Table 2
of this manual.
Use appropriate safety precautions when working with
hazardous liquids and gases.
Be aware of the direction of gas flow to avoid hazardous
conditions during bleed-off procedure.
For pressures above 5 psi and vacuum pressures of -5 psi,
use only securely clamped reinforced tubing.
General Specifications
Power Requirements............................................ One 9-volt battery, NEDA 6LR61,
IEC 1604A
Battery Life ............................................................ 60 hours of continuous use
Operating Temperature........................................ 0 °C to 30 °C (32 °F to 86 °F)
Overpressure ........................................................ Not to exceed 110 psi
Pressure Connection ........................................... Male Luer–Lok
Ranges................................................................... -698 to +802 mmHg
-949 to +1090 cm H
2
O
-374 to +429 in H
2
O
-13.50 to +15.50 psi
-13.5 to +100.0 psi
Accuracy ............................................................... ±1 % of full scale at all ranges
Voltage Output ...................................................... 0.1 V/psi on all ranges except 100 psi,
which is 0.01 V/psi
Voltage Output Accuracy..................................... 5 % of range
Display................................................................... 0.5–in LCD with
LO BATT
Negative polarity indication
Dimensions ........................................................... 14.61 × 9.14 × 4.83 cm
(5.75 × 3.6 × 1.9 in)