19
Front Panel Keys (cont.)
E
Loads a newly keyed-in output value showing on
the Control Display into the Calibrator. If you press
E without identifying the units for the entry,
the Calibrator keeps the units (except Hz) that
were last used. This allows you to alter values for
a function without keying in the units every time.
The multipliers are not retained, for example, you
can enter 1 mV, and then later simply enter 10 to
obtain 10V.
In addition to these functions, the E key
recalls the currently programmed reference value
during error mode operation.
D V A Q H
(Output Function Keys)
These keys select the output function as follows:
D Decibels relative to 1 mW
V Voltage
A Current
Q Resistance
H Frequency
When a value, then H are selected, the
Calibrator automatically switches to ac. To change
back to dc, enter 0 H or enter a signed (+ or -
amplitude).
U m K M
(Multiplier Keys)
These keys select output value multipliers. For
example, if you enter 33, then m, then V, then
ENTER, the Calibrator output value is 33 mV.
U micro (10
-6
or 0.000001)
m milli (10
-3
or 0.001
K kilo (10
3
or 1000)
M Mega(10
6
or 1,000,000)

Other manuals for Fluke II Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke II Series and is the answer not in the manual?

Fluke II Series Specifications

General IconGeneral
True RMSYes
Temperature-40 °C to 400 °C
ContinuityYes
Diode TestYes
Safety RatingCAT III 600V
Voltage DC600V
Voltage AC600V
Current DC10A
Current AC10A
Resistance40MΩ
Battery Life400 hours
Warranty3 years

Related product manuals