Front Panel Keys
O
Toggles the Calibrator between operate and
standby modes. In standby mode, the OUTPUT
binding posts are internally disconnected from the
Calibrator. The Calibrator powers up in standby
mode.
The Calibrator automatically switches to standby
mode when:
r is pressed.
the Calibrator is sourcing an ac or dc voltage
below 22V, and a voltage greater than 22V is
selected.
the output function is changed, e.g., from dc
current to ac current.
the output location is changed.
The Calibrator does not switch to standby when
the voltage is switched from any ac or dc voltage
to an ac or dc voltage less than 22V.