EasyManua.ls Logo

Fluke ScopeMeter 124B - Waveform Acquisition; Single Acquisition

Fluke ScopeMeter 124B
94 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Industrial ScopeMeter®
Scope and Meter Mode
27
Waveform Acquisition
The Test Tool can be setup to acquire waveforms as
needed for a specific application. This section explains
the setup options.
Single Acquisition
To catch single events, you can perform a single shot
(one time screen update).
To set up the Test Tool for a single shot on the input A
waveform:
1. Connect the probe to the signal to be measured.
2. Push to open the Scope and Meter button bar.
3. Push 2 to open the SCOPE SETTINGS menu.
4. Use
YZ to highlight the Update as Single.
5. Push to make the change.
6. Push 4 to exit the menu.
The Test Tool will now show the acquisition screen
and the Information area updates:
Waiting
Test Tool is waiting for a trigger
Run
single acquisition is triggered
Hold
single acquisition is complete
hxv19.eps
To continue with the next single acquisition:
7. Push and wait for another single acquisition
trigger.
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com

Table of Contents

Related product manuals