Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual

 :KHQ\RXDUHILQLVKHGVHWWKH$WR6WDQGE\
Table 4-12. HF AC Voltage Verification Points
UUT
Model
5502A SC... MODE levsine
UUT
Reading B
Voltage Frequency
$OO 9SS 0+] P9WRP9
$OO 9SS 0+] P9WR9
 9SS 0+] !P9
 9SS 0+] !P9
 9SS 0+] !P9
 9SS 0+] !P9
Input C Trigger Sensitivity Test
Note
The test steps for channel C are only for the models 190-104,
190-204, and 190-504.
3URFHHGDVIROORZVWRWHVWWKH,QSXW&WULJJHUVHQVLWLYLW\
 &RQQHFWWKH7HVW7RROWRWKH$DVVKRZQLQ)LJXUH
ADC
B
NORMAL
SCOPE
FLUKE 5502A CALIBRATOR
PM9091
PM9585 - 50 OHM
USE 50 OHM
TERMINATION
4 CHANNELS
For 190-502 and -504
use only TRM50!
SHUIYHUIHSV
Figure 4-9. 5502A Scope Output to Test Tool Input C