Cleaning
Prior to cleaning, remove voltage tester
from all measurement circuits. If the instru-
ment is dirty after daily usage, it is advise-
able clean it by using a damp cloth and a
mild household detergent. Never use acid
detergents or dissolvents for cleaning.
After cleaning, do not use the voltage tes-
ter for a period of approx. 5 hours.
Calibration Interval
The voltage testers must be calibrated pe-
riodically and checked by our service de-
partment at regular intervals to ensure the
specified accuracy of measurement re-
sults. We recommend a calibration interval
of one year.
Battery Replacement
If no signal sound is audible when shorting
the test probe sproceed with the battery re-
placement.
Completely disconnect FLUKE T50 from
the measurement circuit.
Turn the screw (left) in the case (down,
direction of the arrow). Then open and
remove it.
Remove discharge batteries.
Replace with new batteries, type 1.5V IEC
LR03 AAA respecting correct polarity.
• Insert the battery case and close it.
+
-
Fluke T50
Users Manual
11
ГК Атлант Инжинирингофициальный представитель в РФ и СНГ
+7 (495) 109-02-08 sales@bbrc.ru www.bbrc.ru

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke T50 and is the answer not in the manual?

Fluke T50 Specifications

General IconGeneral
BrandFluke
ModelT50
CategoryTest Equipment
LanguageEnglish

Related product manuals