EasyManuals Logo

FMC Technologies Smith Meter T11 User Manual

Default Icon
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
Page 18 • MN01029 Issue/Rev. 0.5 (6/11)
Problem Cause Remedy
Air passing through meter. Check air eliminator.
Calibrator out of adjustment. Readjust calibrator. See MN01028, Pg. 4.
Leakyvalveinproversystem(during Repairprovervalveandreruntest.
test run).
Mechanical friction in meter. Check all meter clearances. See Pgs. 12
through 13. Check bearings for tightness.
Binding or misalignment in driven Inspect accessories. Repair or replace.
accessories.
Loosegeartrainorslippingdrives. Repairorreplace.
Calibrator out of adjustment. Readjust calibrator. See MN01028, Pg. 4.
Oversize clearances in meter. Check meter clearances. See Pg. 12.
Product being eliminated through See below.
vent line.
Air eliminator cylinder seal damaged. Replace cylinder seal.
Air eliminator float ruptured. Replace float.
Foreign particles on cylinder sealing Clean sealing surfaces.
surfaces.
Linkageoutofadjustmentbetween Adjustlinkagecorrectly.
set-stop counter and valve.
Improper setting of “Minimum Setting” See Pushbutton Set-Stop Counter
in set-stop counter. Service Manual.
Mechanical obstruction in valve. Disassemble and remove obstruction.
Inspect valve internals for damage.
Defective pushbutton set-stop See Troubleshooting Section in Push-
counter. button Set-Stop Counter Service Manual.
Plugged orifice. Check orifice in upper piston of P-2 or
PG-2 Valve (pressure system only).
Strainer clogged. Clean strainer. Replace if necessary.
Butterfly valve in air eliminator Repair or replace.
restricting flow.
Back pressure or manual valves do Disassemble and inspect for foreign
not open. particles and/or defects. Repair.
Mechanical obstruction in meter. Disassemble and remove obstruction.
Inspect meter internals for damage.
Pump not operating (pressure Inspect pump for mechanical or elec-
system). trical defects. Replace if necessary.
No product. Supply product.
Air being eliminated from product. Normal - product flow automatically is
restricted during air elimination.
Pump undersized. Replace with pump of sufficient capacity.
Pump not operating properly. Check pump.
Linkageoutofadjustment. Adjustlinkagecorrectly.
Defective main seat seal in valve. Replace seal.
Foreign matter holding main valve Disassemble valve and remove foreign
seat open. matter. Check seal for damage.
Drive coupling defective. Replace.
Internal problems. See Veeder Root Service Manual.
Section 4 – Troubleshooting Table
Over-registration (register-
ing more volume than pass-
ing through meter).
Under-registration (register-
ing less volume than pass-
ing through meter).
Product being eliminated
through vent line.
System fails to shut off
properly after delivering
preset quantity.
Restricted product flow
through meter.
System leaks after shutdown.
Largenumeralcounteror
ticket printer not operating
satisfactorily.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the FMC Technologies Smith Meter T11 and is the answer not in the manual?

FMC Technologies Smith Meter T11 Specifications

General IconGeneral
BrandFMC Technologies
ModelSmith Meter T11
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals