Baseline calibration
Observe:This function is coming very soon. We will inform by a fw and app update.
This function should NEVER replace a calibration of the pumps and pH-electrode,
but is ONLY used if you of some reason have a reagent concentration that is not
correct, thus leading to a significant inaccuracy. The typical, and probably the only
situation, you would need this function is if you mix the reagent yourself, and do
some failure in that process, thus leading to a reagent with not perfect
concentration. Thus this function will save you, and you can still use the reagent. "
Following condition must be true, if this function should be used:"
If you decide to use this function, and all conditions is fulfilled for that (se orange
box), then go to ”Baseline calibration” in settings. "
Now you slide the button to ”ON”, to make it possible to enter data. If you have the
button in OFF-position, it will not be possible to enter any data, and the Baseline
calibration is not active."
In the data-field enter following dKH values:"
"
Field 1: ”dKH value from Alkatronic”
Field 2: ”dKH value from a reference test
Make sure you really have correct values when you enter these in the
data-fields. Thus, concerning the Alkatronic dKH-value, let Alkatronic
measure at least twice and be sure these 2 values are very near each other
as an indication that the measurement is correct performed. And concerning
Correct calibrated pump A and C, and correct
calibrated pH-electrode, but still a dKH measurement
from Alkatronic which is deviating more than 0.2dKH
from the true dKH measured by a high quality
reference test.