7
the measuring beaker, and place the beaker inside the machine in its fitting. "
"
"
Make sure the pipes inside the machine are in
proper position!! The premounted silicon hoses on
the pipes should be in contact with the pipeholder,
and in that way guarantee correct positions, but that
may need to be fine-tuned by the user, so take very
good care to ensure the following pipe-positions (fig
2a,2b):"
Wait to install the pH-probe until it is calibrated, let
it be in the transport cup until then. "
"
Pipe from Pump C should be a few cm above surface of the water.
Pipe from pump B must be in contact with the bottom. Very important!!
Pipe from pump A must be around 2 mm above bottom. Very important!!
Establish software connection
The next step is to establish a connection between the APP in the phone/tablet and
the machine. "
"
General information:
The connection between Alkatronic and App concerning the most crucial
functions, is based on BT-connection, as that is the most safe and stable
connection. But it could take sometimes 5-10 seconds for the app to connect
when opening the app from a sleeping mode. That is normal!. Just wait, and
Figure 2a
Pump A
Pump C
Pump B
Pump D
Figure 2b