In the package we include:"
• Silicon hose, ID=3.2mm."
• Silicon hose, ID=1.6mm"
Please note:"
• Hoses running from tank/bottles should be as vertical as possible. Reduce
as much horizontal length as possible.
• You have to cut the proper length yourself, as that will vary between different
set-ups. "
• The hose for the connections 3,4 and 5, you can use an ordinary 4/6 hose of any
kind, as these hoses are not crucial concerning function or accuracy. "
• But never use other hoses than included concerning connection 1 and 2!
•
It is strongly recommended to have as short length as possible concerning
connections 1(maximum 0.5m) and 2 (maximum 1m, preferably shorter). "
# "
1) Cut the hose 1.6 mm in a proper length (if needed). Connect one end to the inlet
”Reagent In (1)” and the other end to the long plastic pipe. Put this plastic pipe
down in the reagent bottle."
2) Cut the hose 3.2mm in a proper length. Make it as short as possible, and
ideally not longer than 1 meter. Connect one end of hose A to the inlet ”Water In
(2)” and the other end to the included microfilter. Place the microfilter in a sump area
(or other area where you want to take the water sample to measure) where it will
never be exposed above the water´s surface. Never use this hose without the
microfilter, and make sure the microfilter never can float up!
"
3) Cut the 3.2 mm (or similar hose, like an ordinary 4/6 hose
of any kind) in a proper length. Connect one end of this
hose to the outlet ”Waste out(3)”, and place the other end
in a container which will collect the wastewater from the
measuring procedure. If you measure every 6 hours, the
waste volume is around 1.5 liter/week. Every measure
produces around 55 ml of waste water. "
4) Cut the 3.2mm (or similar hose, like an ordinary 4/6 hose
of any kind) in a proper length. Connect one end of this
hose to the inlet ”Alk In (4)” and place the other end in a
container of carbonate-solution, made according to our
recipe (see appendix A). "
5) Cut the hose 3.2mm (or similar hose, like an ordinary
4/6 hose of any kind) in a proper length. Connect one end
of this hose to the outlet ”Alk Out (5)”, and lead that hose
to a sump area with ample circulation."
6) Put the included small white magnetic rod into