30
Make sure you don´t have another unit (phone/tablet) already connected to
Alkatronic. The BT connection can not connect to more than one device at same
time. "
Try to kill the app and open it again. If App after that does not succeed to
connect, and you have BT enabled at phone, and are only a few meters from
machine, then restart the Alkatronic by unplug and plug the power-cable. Then
retry to connect and it should connect. Do not forget to select the action and
wash out modes as these are always o as default when power up (of safety
reason)."
No wifi-connection
The Alkatronic only allows a single connection at a time (Bluetooth via App, or Wifi
to cloud).% When App is connected, the Wifi is disabled and vice versa.%If issues with
connecting to your home router, try this:"
1. Restart the Alkatronic"
2. Wait for BT connection"
3. Clear the wifi credentials and press save"
4. Enter the wifi credentials and press save"
5. Enable alerts for ”dKH high” and ”dKH low”, and press save"
6. Kill app"
7. Restart Alkatronic again. DO NOT OPEN THE APP NOW. "
8. Wait for a few minutes (still do not connect with app)"
9. Check in router to see if you can see “raspberrypi”, if you do, Alkatronic now
communicates with your router."
Once you have got this connection to the router, you will of course not have to do
these actions anymore(and mostly not at all, as also the first connection is
automatically)."
Failsafe message ”Restart”
If LCD shows “Restart”, this is a fail-safe action from Alkatronic due to that the
measured value of some reason was higher than 15.4dKH"
The reason can be one of these:"
1. The true dKh value is really above 15.4 dKH"
2. You have forgot to put the magnetic rod in
the measuring beaker"
3. Stirrer not moves (contact support)"
4. No reagent, or too little reagent is delivered:"
I. Reagent bottle is empty"
II. Reagent pipe not in proper position
in bottle%"
III. Hose inside pump C got