not attaching the plastic housing at all, AND also run in the hose further some
minutes. "
4. Sometimes a newly installed hose, even if it is in perfect position, may stall
a few seconds. This is very soon disappearing after only a few minutes of
running time, and depends on that a completely new hoses is some rigid. "
5. Even if our hoses is of very high consistent, they may have some slight variability
the first days after newly installed. Therefore we advice user to check the
calibration status with ”test pump accuracy” and if necessary recalibrate a few
times the first days. After that, a calibration frequency of 2-4 weeks, should be
sufficient. "
"
Fail-safe messages in LCD
Alkatronic have several fail-safes for security reason. Some of them will show up
with a message in the LCD-display. Here is an explanation of the different fail-safes
messages: "
”Please restart”:
This means that you must restart machine manually, please read about this at page
30. "
“Out of Range/Re-Calibrate”"
This means that when you calibrated the pump/s, you have entered an invalid data,
thus a volume that is out of range. Please then recalibrate that pump. "
Machine will detect out of range if you have entered calibration data outside these
values:"
•
Pump A: 35<X<70"
• Pump C: 5<X<15"
•
Pump D: 5<X<15"