36
attachments, and make sure it is pushed down all the way. "
5. Now push down the hose between the inside wall and wheel.This is more
easy if you simultaneously with your finger rotate the wheel. "
6. Now make sure that the hose is pressed down enough so there is some
space between the rim and hose (see arrow in the picture below).
"
With the larger hose (Pump A,B and D) this could be a little tricky part. Then
use the following method to fine adjust the position of the hose:"
When the hose is in position but just have to be
pressed down a few more mm, start the pump
with ”prime hose”. At same time as the pump
now is running, take the the plastic housing
and carefully replace it. If you feel that the
plastic housing is touching the hose, then just
use the plastic housing to press down the hose
some more, until you feel that the plastic
housing is not touching the hose anymore.
Then the hose is in proper position. If you find
this dicult, contact our support and we will
guide you. "
Important:"
1. When a new hose is installed it requires a short time of run-in, so after the
hose is correctly installed, run the pump with ”prime hose” for around 5 minutes. "
2. Then wait 5 minutes (the hose properties is dierent if it is warmed up by long
time running), and then calibrate the pump."
3. If the pump, after newly installed hose, is stalling, then that depends on that the
hose is to close to the plastic housing. Then try to push it further down so it is