BEFORE TESTING
Calibration
You must calibrate the meter every time you begin to use a new vial of
test strips by setting the meter with the correct code. Test results may
be inaccurate if the code number displayed on the monitor does not
match the number printed on the strip vial.
Step 1
Insert the code strip when the
meter is off. Wait until the code
number appears on the display.
How to Code Your Meter