37
MESSAGE WHAT IT MEAN WHAT TO DO
Appears when the
batteries cannot provide
enough power for a test.
Replace the batteries im-
mediately.
Appears when a used
test strip is inserted
Repeat the test with a new test
strip.
Problem in operation.
Repeat the test with a new test
strip. If the meter still does not
work, please contact the cus-
tomer service for assistance.
Appears when test strip
is removed while count-
ing down, or insufcient
blood volume.
Repeat the test with a new test
strip.
Appears when the
wrong code strip is
inserted or other coding
errors.
Make sure to insert the right
code strip for the meter.
Check if the code numbers on
the code strip and the strip vial
label are the same.
Expired code strip.
Repeat the test with a new
code strip.
Appears when ambient
temperature is below
system operation range
System operation range is
10~40°C (50~104°F).
Repeat the test after the meter
and test strip are in the above
temperature range.
Appears when ambient
temperature is above
system operation
range.
Error Messages