Forma Scientific, Inc.__________________________________________________Start-up
4-2
• Press the ENTER key and the Low Limit temperature along with the word "SET
LOW LIMIT" will be displayed in the LCD window.
• Change the existing Low Limit temperature by pressing the UP or DOWN arrow
keys.
• Press the ENTER key and the LCD display will return to cabinet temperature.
b. To display or change the High Limit Set Point:
• Press the HIGH LIMIT key and a "1" will appear in the LCD window.
• Enter access code (1,2,3,4).
A "1" will remain in the LCD window.
• Press the ENTER key and the existing High Limit along with the word "SET HIGH
LIMIT" will be displayed in the LCD window.
• Change the existing HIGH LIMIT by depressing either the UP or DOWN arrow
keys.
• Press the ENTER key and the LCD display will return to cabinet temperature.
4.3 General Recommendations
Avoid opening the door for extended time periods. Room air, which is higher in
humidity, will replace chamber air and cause frost to develop. Allowing the door to remain
open for extended time periods will cause the chamber to "warm-up", putting undue stress
on the compressors.
This unit is not a "quick-freeze" device. Freezing large quantities of
liquid, or high-water content items, will temporarily increase the
temperature and will cause the compressors to operate for a prolonged time
period. Damage to the compressors may result and product safety may be
jeopardized.
4.4 Preparing the (optional) CoBex Temperature Recorder
a. Installing the chart recorder battery