Forma Scientific, Inc._________________________________________Specifications
8-1
Section 8 - Specifications
8.1
Specification Models 8516 and 8517
Temperature -50° C (-58°F) to -86° C (-123°F)
Exterior
Dimensions
33.25" W x 79.0" H x 3.0" F-B
(85 cm x 201 cm x 94 cm)
Interior
Dimensions
23.0" W x 51.5" H x 25.25" F-B
(58 cm x 131 cm x 64 cm)
Refrigeration Two 1/2 HP Compressors (Cascade System)
Capacity 17.3 cu. ft. (490 liters)
Insulation Type: CFC-free foamed-in-place urethane
Sides: 5" (12.7cm)
Door: 4.5" (11.3 cm)
Electrical 8516: 120 VAC, 1 PH, 60 Hz, 15.5 FLA
Operating Range: 108VAC - 130VAC
8517: 230VAC, 1 PH, 50/60 Hz, 12.0 FLA
Operating Range: 208VAC - 240VAC
Breaker
Requirements
8516: 20 Amp, 120V, Dedicated Circuit, 20 Amp Time Delay Breaker
8517: 15 Amp, 230 VAC
Automatic
Voltage
Compensation
Low:
Cut In: 110V, Cut Out: 115V, Volts Boost: 10
Cut In: 210V, Cut Out: 220V, Volts Boost 18
High:
Cut In: 125V, Cut Out: 120V Volts Buck: 10
Cut In: 235V, Cut Out: 225V, Volts Buck: 18
Shipping Weights Ocean: 1000 lbs. (454 kg)
Air/Container: 900 lbs. (408 kg)
Motor: 758 lbs. (344 kg)