Forma Scientific, Inc.___________________________________________________Service
7-8
Symptom Possible Cause
No Power Light on Monitor. Power line cord disconnected.
Circuit breaker tripped/open.
Main power switch OFF.
Pin-connector not plugged
into the monitor board.
Fuse open on Temp Control board
Main fuse(s) open.
Chamber Temp Deviates from Set
Point
Too much warm product added
Door open too long
Insufficient voltage
Inadequate air circulation
Calibration
Dirty condenser
High Ambient Temperatures
Too Much Frost Build-Up Leaking or damaged door gasket
Freezer not Being Refrigerated
(unit is receiving power)
Compressor thermal overload open
Loss of refrigerant in either system
Defective compressor(s)
Defective temp control
Defective high pressure cut-off
Low stage compressor locked up
Main fuse(s) open
Display Problems in General Defective monitor board